relation: http://publicatio.bibl.u-szeged.hu/31574/ title: Investigation of gamma camera detector uniformity using a dynamic line phantom creator: Séra Teréz Emese creator: Mester János creator: Csernay László creator: Pávics László subject: 03.02. Klinikai orvostan description: BACKGROUND: Investigation of the applicability of the dynamic line phantom for determination o the uniformity of circular and rectangular field-of-view gamma cameras. METHODS: Count rate - activity and uniformity - count rate functions were determined by the dynamic line phantom on three circular field-of-view gamma cameras with 37 PMTs, on one circular field of view camera with 19 PMTs, and on one rectangular field-of view gamma camera with 59 PMTs, with and without a collimator. For an evaluation of the efficacy of the dynamic line phantom, the results were compared with the uniformity values obtained with a 99mTc point source and a 57-Co sheet source. RESULTS: In the optimum count rate range (20,000-30,000 cps for a circular field-of-view, 15,000-65,000 cps for a rectangular field-of-view) determined individually for each gamma camera, the uniformity values obtained with the dynamic line phantom did not differ statistically from the refined with the 99mTc point source or the 57-Co sheet source. CONCLUSIONS: The dynamic line phantom is suitable for the determination of detector uniformity, but the measurements should be performed within a well-defined activity (count rate) interval. date: 1998 type: Folyóiratcikk type: PeerReviewed format: text identifier: http://publicatio.bibl.u-szeged.hu/31574/1/aok_klny_229_98.pdf identifier: Séra Teréz Emese; Mester János; Csernay László; Pávics László: Investigation of gamma camera detector uniformity using a dynamic line phantom. NUCLEAR MEDICINE REVIEW: CENTRAL AND EASTERN EUROPE, 1 (1). pp. 41-45. ISSN 1506-9680 (1998) relation: 2422302 language: eng relation: info:eu-repo/semantics/altIdentifier/doi/ rights: info:eu-repo/semantics/restrictedAccess